ij44c

IJ Series Intermediate Junction Electrodes

Setting the benchmark in super-long life, versatility and value for money since 1972, the flagship of the Ionode Lab sensor range is the renowned IJ Series.

Built to exacting standards and available in pH, ISE, ORP and Reference, the IJ Series benefits are many … A removable sleeve and renewable junction for longer life and easy cleaning, an isolated secondary reference to effectively defeat contamination, an annular ground-glass junction to provide fast response times and resistance to clogging, selectable electrolyte and an optional toughened glass spear membrane for direct penetration. An exceptional choice in virtually any sample, the IJ Series is unequalled in food, wine, dairy, contaminated waste samples, industrial, mining, petro-chemical, electro plating and countless more. The IJ Series also features models with solvent resistant bodies and high temperature housings for the most demanding of environments.

  • Replaceable electrolyte
  • Removable sleeve
  • Available for pH, ISE & ORP
  • Polypropylene body

 

 

 

Way back In 1972,  History of the Intermediate Junction

Back in 1972, Ionode's R&D Chemists and Engineers set out to develop a sensor that would solve the common problems of contamination and clogging etc. but still be tough, fast and super reliable. Working together, they decided it was possible and they set to work on what would become the IJ Series … a sensor that would defeat contamination, work in any sample and have a lifetime far in excess of other single or double junction sensors. The benchmark in sensor versatility, reliability and performance was then born in the IJ Series. Featuring a renewable intermediate junction for easy cleaning, Ionode's proprietary glass formulae and a tough yet flexible body, the IJ Series quickly earned it's reputation for having the longest life in the industry and being the best value sensor available.

 

More Junctions, Less Contamination

IJ have unique junction mechanics which protect the primary reference from the contamination associated with traditional gel electrodes whilst providing the low maintenance and long life of a refillable sensor.

Removable Sleeve

IJ electrodes have a removable sleeve to make cleaning quick and easy.

 

IJ flat

IJ's Explained  Why you need an intermediate junction probe

IJ's have unique junction mechanics which protect the primary reference from the contamination associated with traditional gel electrodes whilst providing the low maintenance and long life of a refillable sensor.  

reference electrode

refillable reference electrodes

 

 

The IJ Range  The full range of IJ electrodes

Swipe to see more information

Model

Parameter

Measuring Range

Temp. Range

Sensor Type

Temperature Compensation

IJ14
Ref
  0 - 60
Reference
n/a
IJ44-Spear
pH
pH 0 - 14
0 - 60
A glass spear
Optional
IJ44-HT-Spear pH pH 0 - 14 0 - 100 A glass spear Optional
IJ40 pH pH 0 - 14 0 - 60 A glass bullet Optional
IJ40-ALK pH pH 0 - 14 0 - 60 C Glass bullet Optional
IJ40-HT-ALK pH pH 0 - 14 0 - 100 C Glass bullet Optional
IJ64
ORP
±2000mV
0 - 60
Platinum Wire
n/a
IJ64D
ORP
±2000mV
0 - 60
Platinum Disc
n/a
IJ64-HT ORP ±2000mV 0 - 100 Platinum Wire n/a
IJAg
Ag+, Cl-
±2000mV
0 - 60
Silver Billet
n/a
IJAu
ORP
±2000mV
0 -60
Gold Wire
n/a
Model Parameter Measuring Range Temp. Range Main Interferences Slope
IJ-F F- 0.2 - 19,000 ppm 0 - 60 OH- 57mV/decade ± 5mV
IJ-Ag2S S2-
0.003 - 32,000 ppm
0 - 60 Hg+/2+ 27mV/decade ± 3mV
Ag+ 0.01 - 108,000 ppm 57mV/decade ± 5mV
IJ-Cl Cl- 1.8 - 35,500 ppm 0 - 60 Br-, I-, S2-, CN-, S2O32- 57mV/decade ± 5mV
IJ-Br Br- 0.2 - 79,900 ppm 0 - 60 I-, S2-, CN- 57mV/decade ± 5mV
IJ-I I- 0.005 - 127,000 ppm 0 - 60 S2-, CN- 57mV/decade ± 5mV
IJ-CN CN 0.2 - 260,000 ppm 0 - 60 S2-, I- 57mV/decade ± 5mV